Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Más filtros










Base de datos
Asunto principal
Intervalo de año de publicación
1.
Biochem Med (Zagreb) ; 34(2): 020802, 2024 Jun 15.
Artículo en Inglés | MEDLINE | ID: mdl-38665873

RESUMEN

We present two cases from the neonatal department with cerebrospinal fluid examination. We revealed a striking discrepancy in polymorphonuclear (PMN) and mononuclear (MN) cell counts using conventional light microscopy in comparison with automated analyzer Sysmex XN-1000 (PMNs - 13 vs. 173x106/L, MNs - 200 vs. 67x106/L in case 1 and PMNs - 13 vs. 372x106/L, MNs - 411 vs. 179x106/L in case 2). We revealed the dominant presence of hemosiderophages in both cases in cytospin slide. Even though Sysmex XN-1000 offers fast examination with a low sample volume, there is possibility of misdiagnosis, with negative impact on the patient.


Asunto(s)
Microscopía , Humanos , Recién Nacido , Microscopía/métodos , Masculino , Femenino , Neutrófilos/citología , Neutrófilos/patología , Líquido Cefalorraquídeo/citología , Recuento de Leucocitos , Leucocitos Mononucleares/patología , Leucocitos Mononucleares/citología
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA
...